Please read these Directions for Use carefully to become thoroughly familiar
with proper operation and maintenance of this bulb suction before using it.
NeoNatalie Suction
Newborn - Reusable
Bulb Suction / Bulb Syringe
Directions for Use
Cat.No. 986000
QTY
1 each
Manufactured in China for:
Laerdal Medical AS, Tanke Svilandsgate 30
P.O. Box 377, 4002 Stavanger Norway
Tel. (+47) 51511700
Fax (+47) 51523557
© 2010, Laerdal Medical AS. All rights reserved.
Distributed in the USA by:
LAERDAL MEDICAL CORPORATION
167 Myers Corners Road, P.O. Box 1840
Wappingers Falls, New York 12590-8840
Tel. (800) 431-1055, +1 (845) 297-7770
Fax (800) 227-1143, +1 (845) 298-4545
20-02076 Rev E
Cleaning After Use
1. Open the Head part and empty remaining suctioned matter.
2. Wash and scrub all parts in soapy water. Flush water through the Nozzle
and Head. Ensure that all visible contamination is removed.
3. Disinfect by one of the following methods, listed in descending order of
efficacy:
Steam autoclaving: 136 °C (277 °F) and 2.0 kg/cm
2
for 10 - 20 minutes
Chemical disinfection: Activated glutaraldehyde solution in accordance
with manufacturer's recommended concentration and procedure. Rinse
in clean water by totally immersing the device for 1 minute.
Boiling in water: Minimum 10 minutes.
4. Dry completely and keep clean until next use.
Technical Specifications
Nozzle dimensions at tip: Inner diameter (ID): 3.0 mm
Outer diameter (OD): 4.5 mm
Suction strength(typical): 100 mmHg [136 cmH
2
O]
Operating temperature: 0 °C (32 °F) to 50 °C (122 °F)
Storage temperature: -20 °C (-4 °F) to 60 °C (140 °F).
Material: Silicone